The new silicone ultra-pure matrix, bonded with octadecylsilane, has better mechanical strength, narrow particle size distribution, excellent batch reproducibility and long lifespan. It has a larger surface bonding coverage and carbon loading, making it an ideal choice for the separation of organic compounds. It is suitable for analyzing complex samples of traditional Chinese medicine.

Hydrogen Bonding Capacity Test Chromatogram
Effective End-capping
- Chromatographic column: SinoPak C18 5μm, 4.6×150 mm
- Mobile phase: Methanol / Water = 30 / 70
- Flow rate: 1.0 mL/min
- Wavelength: 254 nm
- Column temperature: 30°C
- Test compounds: 1. Uracil, 2. Caffeine, 3. Phenol
| Test Item | Evaluation Metric | Detection Result | Conclusion | Reference Criteria |
| Hydrogen bonding capacity | α caffeine/phenol | 0.37 | Low binding capacity | ≤0.6,Low binding capacity>0.6,High binding capacity |
- Chromatographic column: SinoPak C18 5μm, 4.6×150 mm
- Mobile phase: Methanol : 20 mmol/L phosphate buffer (pH 2.7) = 30 : 70
- Flow rate: 1.0 mL/min
- Wavelength: 254 nm
- Column temperature: 30°C
- Test compounds: 1. Uracil, 2. Aniline, 3. Phenol
| Test Item | Evaluation Metric | Detection Result | Conclusion | Reference Criteria |
| Ion exchange sites | α aniline/phenol | 0.06 | Relatively few exchange sites | ≤0.2,Relatively few exchange sites0.2-0.4,Moderate exchange sites>0.4, Many exchange sites |
Excellent Hydrophobic Selectivity and Stereoselectivity
- Chromatographic column: SinoPak C18 5μm, 4.6×150 mm
- Mobile phase: Methanol : Water = 80 : 20
- Flow rate: 1.0 mL/min
- Wavelength: 254 nm
- Column temperature: 30°C
- Test compounds: 1. Uracil, 2. Butylbenzene, 3. Terphenyl, 4. Pentylbenzene, 5. Triphenylene
| Test Item | Evaluation Metric | Detection Result | Conclusion | Reference Criteria |
| Hydrophobic Selectivity | α Pentylbenzene/Butylbenzene | 1.49 | High Selectivity | ≤1.3,Low Selectivity1.3-1.45,Medium Selectivity>1.45,High Selectivity |
| Stereoselectivity | α Triphenylene/o-Terphenyl | 1.69 | High Selectivity | ≤1.3,Low Selectivity1.3-1.45,Medium Selectivity>1.45,High Selectivity |
Excellent Batch Repeatability