The hydrophobic interaction chromatography series includes four models: Amber HIC Ethyl 5μm, Amber HIC Propyl 5μm, Amber HIC Butyl 5μm, and Amber HIC Phenyl 5μm. These columns are specifically designed and developed for high-efficiency separation of proteins, oligonucleotides, and peptides, featuring high resolution, high column efficiency, and excellent reproducibility. The packing matrix consists of uniform spherical, highly cross-linked non-porous polystyrene/divinylbenzene (PS/DVB) material. Ethyl, propyl, butyl, and phenyl hydrophobic ligands are uniformly and densely bonded to the PS/DVB surface, respectively, resulting in four hydrophobic interaction chromatography columns with progressively increasing hydrophobicity. This series is particularly suitable for separating monoclonal antibodies (Mab), antibody-drug conjugates (ADC), and related protein fragments with hydrophobic differences.

  • Monodisperse particle packing, featuring high column efficiency and high resolution
  • Wide pH range: 2 ~ 12
  • Especially suitable for the separation and analysis of monoclonal antibodies (mAbs) and antibody-drug conjugates (ADCs)
  • Maximum operating temperature: 80°C
Amber HIC Butyl
Specification
Stationary PhaseAmber HIC EthylAmber HIC PropylAmber HIC ButylAmber HIC Phenyl
MaterialNon-porous, spherical, highly cross-linked PS/DVB
LigandEthylPropylButylPhenyl
Particle Size1.7μm、5 μm 、 10 μm
pH Stability2-12
MaximumPressure6000、8000(1.7μm)
Maximum Operating Temperature80
Compatible Mobile PhasesConventional aqueous and organic solvents

Application

10 mg/mL Mab 254

  • Column:
  • Amber HIC Ethyl(5μm,4.5×35mm)
  • Amber HIC Propyl(5μm,4.5×35mm)
  • Amber HIC Butyl(5μm,4.5×35mm)
  • Amber HIC Phenyl(5μm,4.5×35mm)
  • Mobile phase:
  • A: 25 mM Sodium phosphate buffer, 2.0 M Ammoniumsulfate, pH 7.0;
  • B: 25 mM Sodium phosphate buffer, pH 7.0
  • Flow rate: 0.8 mL/min
  • Temperature: 25℃
  • Detection: UV280
  • Injection volume: 5 µL

ADC

  • Column: Amber HIC Butyl(5μm,4.6×35mm)
  • Mobile phase:
  • A: 2 M Ammonium sulfate in 0.025 M Sodiumphosphate, pH 7.0,
  • B: 0.025 Msodium Phosphate pH 7.0,C:100%IPA
  • Flow rate: 0.8 mL/min
  • Temperature: 25℃
  • Detection: UV214
  • Injection volume: 10 µL

How It Works

  • Rapid Antibody Titer Determination

Flow rate up to 3 mL/min

Total cycle (including equilibration) time < 0.5 min

  • High Linear Response Range (up to 40 mg/mL)

UV 280 nm: 0.029 – 7.50 mg/mL, LOD 0.294 μg

UV 300 nm: 0.117 – 40.0 mg/mL

  • Durability: > 2000 Injections
  • Reproducibility: Excellent Batch-to-Batch Consistency
  • Suitable for Protein Quantification in CHO Cells with Varying Contents
  • Compatible with HPLC, UPLC and FPLC Systems


The SEM image of Amber Protein A resin shows uniformly distributed particles with a particle size of 20 μm and a particle size distribution of D90/D10 < 1.3.

Ordering Information

NameSize(mm)P/N
Amber HIC Butyl 5μm4.6×353111-81250-021-C