English

Determination of Amoxicillin Sodium and Clavulanate Potassium by HPLC

  • Analyte: Amoxicillin Sodium and Clavulanate Potassium
  • System: EClassical 3200 HPLC with UV Detector
  • Column: Sinopak BEH AQ (5 µm, 4.6×250 mm)
  • Highlight: A reliable HPLC method for amoxicillin sodium and clavulanate potassium meeting ChP requirements with good peak symmetry and resolution
Introduction

Amoxicillin sodium is a broad-spectrum penicillin antibiotic, while clavulanate potassium is a β-lactamase inhibitor that protects amoxicillin from enzymatic degradation. Their combination is widely used in pharmaceutical preparations. According to the 2025 Chinese Pharmacopoeia (draft), the system suitability test for HPLC analysis requires tailing factors between 0.8 and 1.8 for quantitative analysis using peak area. This report presents an HPLC method for the determination of amoxicillin sodium and clavulanate potassium using an Elite 3200 system with UV detection, compliant with pharmacopoeial requirements.

Reagents and Test Sample

Reagents: phosphoric acid (AR), sodium dihydrogen phosphate (AR), methanol (HPLC grade), deionized water (18.2 MΩ·cm), other solvents and chemicals as required for sample pretreatment (e.g., extraction solvents, filtration membranes)

Test sample: the amoxicillin sodium and clavulanate potassium for injection (provided by the client)

System suitability solution: Accurately weigh 0.0084 g of the test sample into a 10 mL volumetric flask, dissolve and dilute to volume with the mobile phase.

Sample solution: Accurately weigh 0.0112 g of the test sample into a 10 mL volumetric flask, dissolve and dilute to volume with purified water.

All solutions are filtered through a 0.45 µm membrane before injection.

Instruments and Equipment

HPLC System: EClassical 3200 equipped with quaternary pump, UV-vis detector, autosampler, column oven, solvent bottle tray, Kromstation CDS.

Pretreatment equipment: analytical balance (0.0001 g sensitivity), ultrasonic cleaner, centrifuge, solvent filtration apparatus, membrane filters, pipettes, etc.

Chromatographic Conditions

Column:              Sinopak BEH AQ (5 µm, 4.6×250 mm)

Mobile Phase:     0.05 mol/L sodium dihydrogen phosphate solution (pH adjusted to 4.0 with phosphoric acid) / methanol = 95 / 5 (v/v)

Flow Rate:          1.0 mL/min

Detection:           UV at 220 nm

Injection Vol.: 20 μL

Column Temp.:   30°C

Experimental Results

System suitability solution separation

A typical chromatogram of the system suitability solution is shown in Figure 1. Four main peaks are well resolved with retention times of approximately 5.845, 7.042, 8.006, and 9.139 min. Chromatographic parameters are summarized in Table 1, all meeting the ChP equirements (tailing factor 0.8–1.8, resolution >1.5).

Fig. 1. Chromatogram of system suitability solution. Peaks, 1-amoxicillin, 2-clavulanate, 3-and 4-impurities

Table 1. Chromatographic parameters of system suitability solution

Peak No. Retention Time (min) Peak Area (mAU·s) Tailing Factor Plate Number (N) Resolution
1 5.845 502.231 1.263 14311
2 7.042 479.725 1.023 11560 5.237
3 8.006 105.851 1.117 12050 3.480
4 9.139 1523.332 1.127 12274 3.647

Method Performance

Method Repeatability

The sample solution was injected ten times consecutively (Figure 2). The retention time RSDs for the four peaks were below 0.33%, indicating excellent system precision. Note that peaks 2 and 3 (retention times ~6.95 and 7.99 min) showed increasing peak areas over time (possibly due to instability), so their area RSDs were not calculated. Detailed data are given in Table 2.

Fig. 2. Overlay chromatogram of 10 replicate sample injections. 

Table 2. Repeatability of sample analysis (10 injections)

Peak No. Retention Time RSD (%) Area RSD (%) Tailing Factor Range Plate Number Range
1 0.15 1.21–1.26 14409–14814
2 0.33 0.91–1.04 6279–7856
3 0.15 1.04–1.07 11360–37878
4 0.20 0.93–0.94 12181–12456
Conclusion

The results demonstrate that the Elite 3200 HPLC system equipped with a Sinopak BEH AQ column meets the system suitability requirements of the Chinese Pharmacopoeia for the analysis of amoxicillin sodium and clavulanate potassium. All peaks exhibit tailing factors within 0.8–1.8 and resolutions >1.5. The method shows good retention time precision (RSD <0.33%).

Share to:
Contact us
×
* Required field
Thanks
Your info had been submitted.